POLYMOBIL III RXB8-115.061.02 Page 2 of 4 Siemens AG
Rev. 01 10.95 CS PS 24 Medical Solutions
1 - 2 Prerequisites
Safety notes and protective measures 1
Connect the POLYMOBIL III only to a power source (power outlet) meeting the
standards of VDE 0107 or the usual local standards.
Before working on the open POLYMOBIL III switch OFF the unit with the power switch at
the control panele and unplug the power cable.
Before removing or inserting PC boards, switch off the generator; in doing so, observe
the ESD regulations.
When performing work and tests, the following must be
observed:
- the product-specific safety notes contained in the document,
- the safety notes RA0-000.012.40... in Register 2 of the
POLYMOBIL III file,
- and the general safety notes contained in Register 2 of the TI
binder.
After the system has been switched off, a dc voltage of approx.
300 V is applied for the main inverter.
On D920 the yellow LED V35 lights up.
Within approx. 1.5 minutes the voltage drops to approx. 12 V;
LED V35 goes out.
Checks and adjustments that must be carried out with radiation
switched on are marked with the symbol
When performing work procedures which are marked with this
symbol, radiation protective clothing must be worn.
CAUTION

Other manuals for Siemens POLYMOBIL III

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens POLYMOBIL III and is the answer not in the manual?

Siemens POLYMOBIL III Specifications

General IconGeneral
BrandSiemens
ModelPOLYMOBIL III
CategoryMedical Equipment
LanguageEnglish

Related product manuals