4 - 1
Siemens AG RXB8-115.061.02 Page 1 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Notes on trouble-shooting 4
In case of errors which impair radiography, perform the following checks.
Checking the line voltage, fuses and LEDs 4
a) Checking the line voltage
Measure the supply voltage on site at the safety socket using the digital multimeter.
Plug the power plug of the POLYMOBIL into the safety socket.
POLYMOBIL ON
POLYMOBIL OFF.
For voltages > 240 V AC see Startup Instructions
RXB8-115.034.02... chapter 3.
If the POLYMOBIL cannot be switched on although the power plug
and power voltage have been connected, the overcurrent circuit
breaker F1 may have responded (automatic circuit breaker
1/Fig. 1)
NOTICE
NOTICE