4 - 1
Siemens AG RXB8-115.061.02 Page 1 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Notes on trouble-shooting 4
In case of errors which impair radiography, perform the following checks.
Checking the line voltage, fuses and LEDs 4
a) Checking the line voltage
• Measure the supply voltage on site at the safety socket using the digital multimeter.
• Plug the power plug of the POLYMOBIL into the safety socket.
• POLYMOBIL ON
• POLYMOBIL OFF.
For voltages > 240 V AC see Startup Instructions
RXB8-115.034.02... chapter 3.
If the POLYMOBIL cannot be switched on although the power plug
and power voltage have been connected, the overcurrent circuit
breaker F1 may have responded (automatic circuit breaker
1/Fig. 1)
NOTICE
NOTICE