POLYMOBIL III RXB8-115.061.02 Page 24 of 24 Siemens AG
Rev. 01 10.95 CS PS 24 Medical Solutions
4 - 24 Notes on trouble-shooting
Checking the control elements on the collimator 4
Check the light-beam localizer operated both at the control panel and at the collimator.
To do so, press the button (6/Fig. 10 or 7/Fig. 9),
the light-beam localizer lamp lights up for 20 s
± 1s.
Check if the double-slot diaphragm can be rotated ± 45°.
Check that format adjustment (blades) is easy and smooth by turning the collimator
adjustment knobs (8/Fig. 9).
6
Fig. 9 Fig. 10
7
8