4 - 8 Notes on trouble-shooting
Checking the intermediate circuit voltage 4
• Connect oscilloscope or digital multimeter to D111.R1 and R2.
• Depending on the supply voltage, a value between 250... 350 V DC must be measured
with the digital multimeter.
Charging and discharging procedure
On D111 an intermediate circuit voltage of ... up to 350 V is
applied!
During oscilloscope operation the protective ground wire connec-
tion in the power cable must not be interrupted under any circum-
stances.
For measurements where ground loops that may be present could
impair the measuring result, use the TEK amplifier and the trigger
attachment.