Notes on trouble-shooting 4 - 7
Siemens AG RXB8-115.061.02 Page 7 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
• Measure the supply voltages on D910.X3.
+ 5 V at X3.4 and X3.1 (0 V) Tolerance
± 0.5 V
+15 V at X3.5 and X3.6 (0 V) Tolerance
± 1 V
- 15 V at X3.7 and X3.6 (0 V) Tolerance
± 1 V
D910 (X038I-91E)