POLYMOBIL III RXB8-115.061.02 Page 10 of 24 Siemens AG
Rev. 01 10.95 CS PS 24 Medical Solutions
4 - 10 Notes on trouble-shooting
Oscillating current measurement 4
• POLYMOBIL OFF
• Connect oscilloscope on D920 to MP "I
s
" and "GND".
• POLYMOBIL ON
• Trigger exposure with default values 60 kV, 10 mAs.
Main inverter ok
• During the kV rise
• After the kV rise
Integrated current transformer.
1V 50 A
NOTICE
1 V 50 A
1 V 50 A