5 - 1
Siemens AG RXB8-115.061.02 Page 1 of 6 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Replacing important components 5
Replacing the single tank 5
To replace the single tank, remove it from the mobile stand as follows:
Move the arm system in the lowest position and check if the safety locking (pin 1/Fig. 1)
has engaged.
The stand cannot be moved upwards any longer.
Tighten the turning knob(3/Fig. 1).
Loosen the threaded ring and pull out the connector of the single tank.
Using an 8 mm Allen key, loosen the two screws on the bracket (2/Fig. 1) approx. two
turns.
Lift the single tank off vertically and place it on a soft base.
Detach the collimator (see chapter 5, page 2).
Install the new single tank in the reverse order.
Check alignment of light field to radiation field and adjust if necessary. (see "Aligning the
light field to the radiation field").
Fig. 1
1
2
3