5 - 1
Siemens AG RXB8-115.061.02 Page 1 of 6 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Replacing important components 5
Replacing the single tank 5
To replace the single tank, remove it from the mobile stand as follows:
• Move the arm system in the lowest position and check if the safety locking (pin 1/Fig. 1)
has engaged.
The stand cannot be moved upwards any longer.
• Tighten the turning knob(3/Fig. 1).
• Loosen the threaded ring and pull out the connector of the single tank.
• Using an 8 mm Allen key, loosen the two screws on the bracket (2/Fig. 1) approx. two
turns.
• Lift the single tank off vertically and place it on a soft base.
• Detach the collimator (see chapter 5, page 2).
• Install the new single tank in the reverse order.
• Check alignment of light field to radiation field and adjust if necessary. (see "Aligning the
light field to the radiation field").
Fig. 1
1
2
3