Notes on trouble-shooting 4 - 15
Siemens AG RXB8-115.061.02 Page 15 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Setting the maximum filament frequency 4
POLYMOBIL OFF
Remove fuse F5 on D920.
Connect oscilloscope on D910 to MP "CAL" and "GND".
POLYMOBIL ON
Using P3, set the maximum filament frequency on D910 (100 kHz 10 µs)
Err 3 is displayed.
NOTICE