Notes on trouble-shooting 4 - 15
Siemens AG RXB8-115.061.02 Page 15 of 24 POLYMOBIL III
Medical Solutions Rev. 01 10.95 CS PS 24
Setting the maximum filament frequency 4
• POLYMOBIL OFF
• Remove fuse F5 on D920.
• Connect oscilloscope on D910 to MP "CAL" and "GND".
• POLYMOBIL ON
• Using P3, set the maximum filament frequency on D910 (100 kHz 10 µs)
Err 3 is displayed.
NOTICE