POLYMOBIL III RXB8-115.061.02 Page 4 of 6 Siemens AG
Rev. 01 10.95 CS PS 24 Medical Solutions
5 - 4 Replacing important components
• Check function of the light-beam localizer lamp.
The light-beam localizer lamp voltage can be adjusted at resis-
tance R6 (12/Fig. 6).
The voltage at the lamp should be 11.5 V
± 0.2 V.
12
Fig. 6
NOTICE