POLYMOBIL III RXB8-115.061.02 Page 4 of 24 Siemens AG
Rev. 01 10.95 CS PS 24 Medical Solutions
4 - 4 Notes on trouble-shooting
c) Checking the LEDs on D910 and D920
POLYMOBIL ON
After the initialization, the default data 60 kV and 10 mAs are displayed.
Stand-by operation
On D920 the following LEDs are illuminated:
With a supply voltage of 110 V, LED V12 (green) is illuminated on
D910.
V40 (yellow) + 24 V
V39 (yellow) V
filament
V41 (yellow) I / 0 (ON/OFF)
V35 (yellow) + 300 V intermediate circuit voltage
D920 (X038I-92D)