Maintaining the System Page 3-25
Replacing the Probe and Tubing and Probe Housing
6. Carefully hold the probe sleeve and pull firmly to remove the probe
housing. Release the probe lever.
Figure 3-30. Removing the Probe Housing
7. If you are replacing both the probe and the housing go to step 12. If
not continue with step 8.
8. If you want to clean the probe and housing:
Soak the assembly for 10 minutes in 10% v/v bleach.
Rinse with deionized water and gently dry with tissues.
Lightly grease the probe shaft mechanism (use the grease
supplied in the Spares box).
Figure 3-31. Greasing the Probe Shaft Mechanism
9. Disconnect the probe tubing from the housing. Pull the probe out of
the housing. Discard the old part.
WARNING Make sure the old probe is disposed of safely, in accordance
with your laboratory guidelines.
Probe Shaft Mechanism