SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 115
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
800dbxxx
800dcxxx
etCO2 Pod/Cartridge - related Try swapping out airway
adaptor with known-good
components.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
800ddxxx
800dexxx
HEMOPOD / HEMOMED - related Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
800dfxxx
800e0xxx
Pod communications
problem.
Pod cable, Pod, or
CPU board failed
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
800e1xxx
800e2xxx
NIBP - related Try disconnecting/
reconnecting pneumatic
tubing.
If problem persists, replace
Main Processor Board
Subassembly.
800e4001 Remote View and
Wedge reset
Upgrade SW to ≥VC2.x.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action