Service Manual SC 7000 and SC 9000XL Patient Monitors
116 Siemens Medical Systems, EM-PCS, Danvers ASK-T898-03-7600
NOT A CONTROLLED DOCUMENT
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
800e5xxx
800e6xxx
RESP-related Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
800e7xxx SPO2 - related Try disconnecting/
reconnecting cable.
If problem persists, try
replacing sensor.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
800e9xxx
800eaxxx
Temperature - related (MultiMed Pod) Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
800f1xxx
800f2xxx
Temperature - related (MultiMed Pod) Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action