SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 117
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
800f5xxx
800f6xxx
HEMOPOD / HEMOMED - related Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
800fbxxx
800fcxxx
NIBP - related Try disconnecting/
reconnecting pneumatic
tubing.
If problem persists, replace
Main Processor Board
Subassembly.
80101xxx
80102xxx
Analog Out Replace Main Processor
Board Subassembly.
80107xxx
80108xxx
RESP-related Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
8010dxxx
8010exxx
SPO2 - related Try disconnecting/
reconnecting cable.
If problem persists, try
replacing sensor.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action