SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 119
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
8016a00a Image checksum test failed due to TAXI
error!
Bad memory/access/
fetch
Restart monitor. If error
persists replace monitor.
80179xxx
8017axxx
Pod communications
problem.
Pod cable, Pod, or
CPU board failed
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
8017fxxx
80180xxx
Temperature -related (MultiMed Pod) Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
8018301f Unable to update CPS Chassis with
pick_n_go_mode
Write failed to CPS or
IDS chassis
EEPROMs
replace CPS or IDS
80185008 Pick and Go reset Upgrade SW to ≥VC1.x.
80187020 Error Log was cleared! Manual clear of error
log via system
console menu
None
8018bxxx
8018cxxx
etCO2 Pod/Cartridge - related Try swapping out airway
adaptor with known-good
components.
Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly.
8018d005 sn A, status B etCO
2
Hardware Fail Replace etCO
2
cartridge.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action