SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 123
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
8023102c LCOM_ROM_DIAG_ERROR Net comm failure If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
8023102e LCOM_SHRAM_CHECKSUM_ERROR Net comm failure If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
80231030 LCOM_SHRAM_DIAG_ERROR Net comm failure If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
80231031 LCOM_SHRAM_PACKET_ERROR Net comm failure If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
80231032 LCOM_SHRAM_PRL_ERROR net comm failure If this message is
accompanied by another
error, check Docking station
connections. Otherwise this
indicates a SW error and
should be reported to
Customer service via
DataCard.
8023dxxx
8023exxx
80249xxx
8024axxx
PCMCIA Interface - related Retry download. If error
persists, return card.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action