SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 125
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
80271xxx
80272xxx
Network Communications - related Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
80273xxx
80274xxx
NIBP - related Try disconnecting/
reconnecting pneumatic
tubing.
If problem persists, replace
Main Processor Board
Subassembly.
8027dxxx
8027exxx
Network Communications - related Try disconnecting/
reconnecting cable(s)
between Monitor and
network wallbox.
Verify proper SW version(s)
installed in all related
devices, in accordance with
SW compatibility chart for
Monitor SW.
If problem persists, try
swapping CPS/IDS with
known-good device.
If problem persists, replace
Main Processor Board
Subassembly.
8027fxxx
80280xxx
Software did not
allocate enough
memory for user’s
current setup.
Try disabling items
connected or disabling Auto
Dual View.
If error persists, replace
Main Processor Board
Subassembly.
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action