SC 7000 and SC 9000XL Patient Monitors Service Manual
ASK-T898-03-7600 Siemens Medical Systems, EM-PCS Danvers 127
7k9kXLSM.apc.CD_ROM.fm/04-99/kaupp
NOT A CONTROLLED DOCUMENT
802abxxx
802acxxx
802adxxx
802aexxx
RECORDINGS - related Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping Recorder.
If problem persists, replace
Main Processor Board
Subassembly.
802b9xxx
802baxxx
RESP-related Try disconnecting/
reconnecting cable.
If problem persists, swap
cable with known-good
cable.
If problem persists try
swapping pod.
If problem persists, replace
Main Processor Board
Subassembly
802bbxxx
802bcxxx
Try upgrading SW to
currently released version.
If error persists, replace
Main Processor Board
Subassembly.
802bf008 memory partition 2 is 91 percent used. Memory utilization
note
Ignore
802d4002 shutdown not due to user. Could be low
battery.
Monitor was shut
down without using
the user’s front panel
switch. Usually this
means that the
monitor was running
on internal batteries
and ran out of power.
If runing monitor on internal
battery, ignore this
message. Otherwise report
condition to customer
service
Table C-1Diagnostic Log Codes (Continued)
Code
xxx = any
alphanumeric
characters
Description Cause Action