Setup with the integrated HMI
7.3Menu overview
Menu item Parameter Setting Default
AFES OFF False Signal Calculation: Manual
Curve not active
AFES
RANGE
False Signal Manual Action:
Distance
0mm
MIN DISTANCE Min Level Distance -
MAX DISTANCE Max Level Distance -
TR TEMP MIN Min Electronic Temperature -
PEAK VALUES
TR TEMP MAX Max Electronic Temperature -
LOOP TEST LOOP TEST Current Output Simulation OFF
CANCL No restartDEVICE RSTRT =
Warmstart
OK Make restart
CANCL
NO No Reset
FACT Basic settings
RESET
RSTR
CUST
CFG
Factory settings
NO
NO No Reset
DIST Reset User Min Max Level
Distance. Reset Action = Both
values
RESETS
RESET PEAK
STEMP Reset User Min Max Electronic
Temperature. Reset Action =
Both values
NO
FREQUENCY 1: EU, Albania, Andorra,
Azerbaijan, Australia,
Belarus, Bosnia and
Herzegovina, Canada,
Iceland, United Kingdom,
Liechtenstein, Moldova,
Monaco, Montenegro,
New Zealand, Northern
Macedonia, Norway, San
Marino, Saudi Arabia,
Switzerland, Serbia,
Turkey, Ukraine, USA
2: South Korea, Taiwan,
Thailand
3: South Africa
4: Russia
Radar Mode country selection 1
Menu item Parameter Setting Default
POLLING ADDR Display Read Write Hart Parameter, Polling
Address
0
Menu item Parameter Setting Default
USER PIN ACTIVATE, CANCEL ENABLE
BLUETOOTH PIN Bluetooth Access Code Individual
SITRANS LR150
Operating Instructions, 08/2020, A5E50352155
33