Setup with the integrated HMI
7.5.2Quick setup level, space, distance
7.5.2 Quick setup level, space, distance
Mode
OPERATION: LEVEL, SPACE, DIST
Sets the mode of operation that determines the output and display:
Setting Description
LEVEL Level
SPACE Space
DIST Distance
VOL Volume
VFLOW Volume flow
1
Distance
2
Space
3
Upper calibration point
4
Transmitter reference point
5
Lower calibration point
6
Level
Figure7.6 Overview quick setup level, ullage, distance, user-specific
Setting Description Reference point
LEVEL Material height Lower calibration point (zero point of
the process)
SPACE Distance from the Upper calibration
point to the material surface
Upper calibration point (full point of the
process)
DIST Distance to the material surface Lower calibration point
36
SITRANS LR150
Operating Instructions, 08/2020, A5E50352155