4BWEIGHING FUNCTIONS
Bit Meaning
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Char. value of the load cell
0 0 ≤ 1 mV/V
0 1 ≤ 2 mV/V (default)
1 0 ≤ 4 mV/V
1 1 Reserve
Decimal place for remote display
0 0 0 Decimal place 0 (default)
0 0 1 Decimal place 1
0 1 0 Decimal place 2
0 1 1 Decimal place 3
1 0 0 Decimal place 4
1 0 1 Decimal place 5
Filter setting
0 0 0 No filtering
0 0 1 Cutoff frequency: 5 Hz
0 1 0 Cutoff frequency: 2 Hz (default)
0 1 1 Cutoff frequency: 1 Hz
1 0 0 Cutoff frequency: 0.5 Hz
1 0 1 Cutoff frequency: 0.2 Hz
1 1 0 Cutoff frequency: 0.1 Hz
1 1 1 Cutoff frequency: 0.05 Hz
Mean value filter
0 No mean value filter (default)
1 Mean value filter switched on (30x20ms)
Memory goal
(only relevant for limit values and zero set value via
set to zero command)
0 Carry over into EEPROM (default)
1 Keep in RAM only
Operational state - channel
0 Channel switched on (default)
1 Channel switched off (default channel2 of U1)
Reserve
0
1
Weight unit
(only relevant for SIWATOOL, is only saved in SIWAREX U)
0 0 0 ‚ ‚(default)
0 0 1
‚ g ‚
0 1 0
‚ kg ‚
0 1 1
‚ t ‚
1 0 0
‚ oz ‚
1 0 1
‚ lb ‚
1 1 0
, lbs‚
1 1 1
‚ kN ‚
Reserve
(Reserve, currently not used)
0 (default)
1
Figure 5-2 Parameter table for setting data
SIWAREX U 5-29