4BWEIGHING FUNCTIONS
0 no display
1 4-digit display Siebert display
2 5-digit display Siebert display
3 6-digit display Siebert display
Assignment LED1
BYTE 1 101
Code Status information Channel no.
100 Channel error Channel 1
101 Limit value 1 Channel 1
102 Limit value 2 Channel 1
103 Scale adjusted Channel 1
200 Channel error Channel 2
201 Limit value 1 Channel 2
202 Limit value 2 Channel 2
203 Scale adjusted Channel 2
100 Channel error Channel 1
Assignment LED2 BYTE 1 102 Reserve
Reserve BYTE 1 0
6
Table 5-2 Allocation of DR 5
5.4.1 DR 5 – module number
Each module can be assigned a number. With SIWAREX U modules with order
number 7MH4601… and when using SIWATOOL U Version 1.7, the number of the
module is checked against the number saved in the SIWAREX module while
establishing an online connection. Establishing a connection is only possible when
the number in SIWATOOL U matches that of the module or when the number "0" is
specified.
This module number only serves information purposes from SIWATOOL U Version
2.0 on or for modules with order number 7MH4950…. The online connection is
established in every case.
5.4.2 DR 5 – port parameters
The setting of port parameters is only possible to a limited degree. This setting is
valid both for the RS232 port and the TTY port. The baud rate is fixed at 9600 bit/s.
Bit no. reset set
0 with parity bit (default) without parity bit
1 parity even (default) parity odd
2 - 7 reserve
5.4.3 DR 5 – display model
The Siebert remote display Model S102 can be connected to the SIWAREX U
module via the TTY port. Dependent on the number of display elements,
corresponding codes are allocated.
Code (dec) Meaning
5-34
SIWAREX U