12BTECHNICAL DATA
Attention
This is a class A device. In residential areas, this device can cause radio
interference. In such areas, users must take suitable measures (e.g.: use in 8MC
cabinets)
.
13.10.3 Climatical requirements
Climatical requirements
Comments Environmental conditions Areas of use
Operating temperature:
horizontal installation
other installation positions
DIN EN 60068-2-2
0 to +60°C
0 to +40°C
Storage and transport
temperature
DIN EN 60068-2-2
- 40 to +70°C
Relative air humidity 15 to 95% Without condensation, relative
humidity (RH) corresponds to
stress factor 2 in accordance with
DIN IEC 61131-2
Air pressure in operation DIN EN 60068-2-13
from 1080 to 795 hPa
Corresponds to an altitude
of -1000 to 2000 m above sea level
Air pressure in storage DIN EN 60068-2-13
from 1080 to 600 hPa
Corresponds to an altitude
of -1000 to 3500 m above sea level
Function-endangering gases
DIN IEC 60721-3-3
Class 3C3
SO
2
: 10cm
3
/m
3
H
2
S: 1cm
3
/m
3
Function-endangering dust
DIN IEC 60721-3-3
Class 3S2
without occurrence of sand
Table 13-12 Data: Climatical requirements