Errors and messages
9.6 Operating errors
SIWAREX WP251
168 Manual, 12/2015, A5E37203357A
Data and command errors
Error code
Additional
string
Description
4519
, because no test operation (only relevant in test field)
4522 , because calibration not yet complete
6053 Calibration command
error
6053 Command from the Calibration commands group could not be
executed. Additional information contains details.
4510
, because service mode is not active
4511
, because write protection active
4520
, because increment between calibration digits is too small
4521
, because order of calibration points is incorrect
4522
, because calibration not yet complete
4523
, because calibration digits not in the permissible range
4524
, because calibration weight 0
4527
, because command is not permitted in simulation mode
6054 Scale command error 6054 Command from the Scale commands group (Zero, Tare, etc.)
could not be executed. Additional information contains details.
4500
, because a fault is present
4501
, because no standstill occurred
4502
, because module is already waiting for standstill
4530
, because weight is outside the permissible tare weight range
4531
, because weight is outside the permissible zero setting range
6055 Weighing command
error
6055 Command from the Weighing commands group (Start single
weighing, Start continuous mode, etc.) could not be executed.
Additional information contains details.
4500
, because a fault is present
4501
, because no standstill occurred
4502
, because module is already waiting for standstill
4508
, because dosing is not active
4509
, because timing within dosing step sequence is impermissible
4513
, because service mode is activated
4515
, because not possible in this operating mode
4540
, because weight above fine shut-off point
4542
, because set point too small
4543
, because set point too large
4544
, because tolerance limits implausible
4545 , because fine weight, trailing weight and set point are incompati-
ble
4546
, because dosing would cause overfilling
4547
, because configured emptying implausible
4548
, because no check stop points defined
4549
, because scale is already empty