Errors and messages
9.6 Operating errors
SIWAREX WP251
168 Manual, 12/2015, A5E37203357A
, because no test operation (only relevant in test field)
4522 , because calibration not yet complete
6053 Calibration command
error
6053 Command from the Calibration commands group could not be
executed. Additional information contains details.
, because service mode is not active
, because write protection active
, because increment between calibration digits is too small
, because order of calibration points is incorrect
, because calibration not yet complete
, because calibration digits not in the permissible range
, because calibration weight 0
, because command is not permitted in simulation mode
6054 Scale command error 6054 Command from the Scale commands group (Zero, Tare, etc.)
could not be executed. Additional information contains details.
, because a fault is present
, because no standstill occurred
, because module is already waiting for standstill
, because weight is outside the permissible tare weight range
, because weight is outside the permissible zero setting range
6055 Weighing command
error
6055 Command from the Weighing commands group (Start single
weighing, Start continuous mode, etc.) could not be executed.
Additional information contains details.
, because a fault is present
, because no standstill occurred
, because module is already waiting for standstill
, because dosing is not active
, because timing within dosing step sequence is impermissible
, because service mode is activated
, because not possible in this operating mode
, because weight above fine shut-off point
, because set point too small
, because set point too large
, because tolerance limits implausible
4545 , because fine weight, trailing weight and set point are incompati-
, because dosing would cause overfilling
, because configured emptying implausible
, because no check stop points defined
, because scale is already empty