Operation requiring official calibration
12.2 Verification
SIWAREX WP251
Manual, 12/2015, A5E37203357A
201
4. Check of the logbook entries
Only the software downloads for the legal trade display SecureDisplay are retained in th
e
l
ogbook. It is no longer possible to download the firmware for the SIWAREX WP2
51
ev
aluation electronics following verification of the scale with the jumper for writ
e
pr
otection inserted.
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
Select the "Logbook check" submenu in the "Calibration check" submenu.
Navigate in the logbook entries using 4 commands:
Display first logbook entry
Display last logbook entry
Display previous logbook entry
Display next logbook entry
5. Checking of the ID labels
Checking is carried out in accordance with the specifications in the EC construction
license.
6. Checking of the serial number of the evaluation electronics
The serial number of the evaluation electronics used (shown at top right in t
he
S
ecureDisplay) must agree with the ID label. The serial number of the evaluati
on
el
ectronics used can also be called on the SecureDisplay.
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
Select the "Version and variables check" submenu in the "Calibration check"
submenu.
Activate the "Display SIWAREX serial number" command.
The current serial number is then shown on the display.
7. Checking of the regulation code
Check whether the regulation code "OIML" is set in the scale parameters. The code c
an
be c
alled on the SecureDisplay.
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
Select the "Version and variables check" submenu in the "Calibration check"
submenu.
Activate the "Display calibration regulation" command.
The currently set code is then shown on the display.