Operation requiring official calibration
12.2 Verification
SIWAREX WP251
202 Manual, 12/2015, A5E37203357A
8. Checking of the additive tare range
When using the additive tare compensation unit, you must check the complete weighi
ng
r
ange (up to maximum load + additive maximum tare load). To do this, you must check
up to the maximum load, and repeat this following taring. Repeat these steps until th
e
upper
limit of the range of the additive tare compensation unit has been reached.
Call up the main menu in the main display (function key with the open-end wrench).
Select the "Calibration check" submenu in the main menu.
Select the "Scale check" submenu in the "Calibration check" submenu.
Check with the "Tare setting" or "Delete tare" command.
9. Checking of the parameter disabling/calibration bridge
The calibration bridge must be inserted on the weighing module. You can check t
he
c
alibration bridge using the specific symbol
at the bottom right in the main display
(symbol displayed = calibration bridge inserted).
10.When using the second display S102 - checking of set telegram address
The telegram address (No. 01) must be set on the Siebert display S102. Checking is
carried out in accordance with the specifications in the Instruction Manual "Siebert series
S102 - digital displays for Siemens Siwarex".

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens SIWAREX WP251 and is the answer not in the manual?

Siemens SIWAREX WP251 Specifications

General IconGeneral
BrandSiemens
ModelSIWAREX WP251
CategoryAccessories
LanguageEnglish

Related product manuals