41 63 189 D 3311
D 3311.201.01.10.02
3
Sirona Dental Systems GmbH
I
nterference with electromedical devices by radio telephones
To warrant the operational safety of electromedical devices, it is recommended that the
operation of mobile radio telephones
in the medical practice or hospital
is prohibited
.
Waste disposal
Public law may include special regulations regarding the waste disposal of this product. To prevent environmental damages
and personal injuries, we therefore ask you to contact us prior to discontinuing the use of this product, with the aim of disposal.
Patient chair
Please observe the maximum load capacity of the chair of
max. 135 kg!
Technical Description
Model denomination:
C2
Power supply: 230VAC 50Hz115VAC 50/60Hz
Total current consumption: 4.5A at 230V 9.5A at 115V
Main fuse for 230VAC: slow blow T6.3 A H, 250VAC, Order No. 10 77 452
Main fuse for 115VAC: slow blow T 10 A H , 250VAC, Order No. 10 77 460
Class
I
equipment
Protection against electric shock:
Type B
equipment
Protection against harmful ingress penetration of water: ordinary equipment (without protection)
Footswitch protected against dropping water IPX 1.
Mode of operation: continuous operation with suspension strain according to dental working procedures.
Permanently connected unit.
This product is provided with a
CE
marking in accordance with the regulations stated in the Directive 93/
42/EEC of June 14, 1993 concerning medical products.
0123