12 Basic Workstation Settings with Mode Key Sirona Dental Systems GmbH
41 63 189 D 3311
34
D 3311.201.01.10.02
12.8 Mode: CUP FILL ON S
YES
-
NO
Filling of tumbler in chair program S
The tumbler is filled when the treatment
chair is in mouth rinsing position
S
(See
page 19).
12.9 Mode: MAX LIGHT VOLT
Instrument light, voltage setting for
instruments.
First a message will appear prompting you
to
TAKE
an
INSTRUMENT
.
Then the current voltage setting of the
withdrawn instrument is displayed.
Using the
–
/
+
keys, you can then set the
instrument light voltage from 3.0 to 5.6 V in
steps of 0.1 V.
Starting from 3.9 V,
WARNING
appears
on the display.
After the instrument has been returned to
its hanger, please set the next instrument.
Do not continue on in the Mode menu
before the instrument light voltage has
been set for all instruments.
12.10 Mode: MAX DRIVE INTEN.
For SIRONA motor and turbine
First a message will appear prompting you
to remove an instrument.
Then the instrument must be started.
The maximum permissible motor speed in
RPM
for the SIRONA motor and the
maximum permissible air pressure in
bar
(or
PSI
) for the turbine can be set with
the
–
/
+
keys.
Once an instrument has been returned to
its hanger, please set the next instrument.
Do not continue on in the Mode menu
before the maximum power has been set
for all instruments.
CUP FILL ON S
YES
−
NO
A
TAKE INSTRUMENT
A
MAX. LIGHT VOLT
3.9 V WARNING
A
TAKE INSTRUMENT
A
START INSTRUMENT
A
A
MAX DRIVE INTEN.
RPM
A
bar