10 Instrument Functions Sirona Dental Systems GmbH
41 63 189 D 3311
24
D 3311.201.01.10.02
10.3 SIROSON L
Once the
SIROSON L
handpiece has been
removed and started, the programmed
intensity value in
%
and the operator
A
or B
appear on the display.
The intensity is set with the
–
/
+
keys from
1% to 100%.
If you
hold
the corresponding key down,
the intensity is adjusted in counting steps:
1%: 1 – 4%
5%: 10% – 100%
Fast adjustment:
If you
briefly touch
the relevant key, the
speed preselection always jumps to the next
step of 20% up or down.
The cooling water flow is always switched
on automatically.
10.4 SIROSON L
−
endodonty
The intensity range of 1-4% is provided for
endodontics. It is adjustable in steps of 1%.
The warning
ENDO MAX. 4%
appears on
the display in this case
For reasons of security, we recommend
that you program the set value for endo-
donty. Otherwise, the value programmed
until now becomes active again during
storing the instrument in the carrier in the
meantime.
ATTENTION
If 4% is exceeded during the
endodonty treatment, temperature dam-
ages and needle breaks can arise.
A
%
ENDO MAX. 4%
A
%