SEMI-AUTOMATIC
10
RF (radiofrequency) interference from devices such as mobile phones and radio two-way transmitters, can cause
SMARTY Saver Plus to malfunction. SMARTY Saver Plus must be kept at least 2 metres away from such RF devices, as
indicated in the IEC/EN 61000-4-3 standards. Keep at sufficient distance from other therapeutic and diagnostic sources
of energy (e.g. diathermy, high-frequency surgery, magnetic tomography).
Use SMARTY Saver Plus only if you have followed a BLS-AED or ALS-AED training course.
Before using the device, make sure that it is not obviously damaged.
Do not use universal defibrillation pads SMT-C2001 in paediatric mode on adult patients (older than 8 and weighing
more than 25 kg). In fact, in paediatric mode, SMARTY Saver Plus automatically reduces the maximum energy that can
be delivered to 50J.
Place the patient cables in such a way as to reduce the possibility of them getting entangled in or strangling the patient.
In a domestic environment, keep the defibrillator away from the reach of children and pets.
Disconnect the patient from equipment that is sensitive to high voltage pulses, or equipment that is not defibrillator-
proof, before delivering the shock.
WARNING
Do not allow the defibrillation electrodes to touch or to come into contact with ECG electrodes, pads, transdermal
plasters, etc. Otherwise, the formation of electric arcs and burns to the patient could be caused during defibrillation;
the current may even be dispersed.
Place the defibrillation pads as indicated in this user manual and on the packaging.
Do not use the defibrillation pads if the gel has come away from the support or if it appears torn, divided or dry.
If damage has been found, in no case should you switch SMARTY Saver Plus on.
Before using the device, remove metal objects from the patient’s body (including necklaces or bracelets, etc.).
Do not use defibrillation pads other than those supplied by the manufacturer. Otherwise, the defibrillator may perform
false interpretations.
Do not use the defibrillation pads if they are damaged, even partly.
Do not touch the patient or the defibrillation pads during the automatic analysis of the heartbeat.
Moving or transporting the patient during the analysis of the heartbeat by the device may lead to a wrong or untimely
diagnosis. Reduce movements to a minimum while the heartbeat is being analysed. If the device is used in a moving
ambulance, stop the vehicle and only start driving after the shock has been delivered.
In order to use SMARTY Saver Plus, you must have followed a basic or advanced training course for cardiopulmonary
resuscitation with use of defibrillator (BLS-AED or ALS-AED course).
Avoid using the universal defibrillation pads SMT-C2001 in adult mode on children (aged 1-8 or weighing 8-25 kg). In
fact, in adult mode, SMARTY Saver does not automatically reduce the maximum energy that can be delivered to 50J and
may, therefore, become hazardous for the paediatric patient.
Avoid using the universal defibrillation pads SMT-C2001 in adult mode on children (aged 1-8 or weighing 8-25 kg).
When paediatric mode is set, the maximum energy that can be delivered is 50J.
If needed, before applying the defibrillation pads dry the patient’s chest and remove excess hair.
Do not subject SMARTY Saver Plus, its accessories and/or its parts to falls and/or strong impacts.
Do not use damaged accessories and/or parts; otherwise, the device may be caused to malfunction.
Use solely original accessories and/or spare parts.
Avoid handling the device, its accessories or its parts too aggressively to avoid possible damage. Inspect the entire
system regularly.
Sanitise the device in compliance with the rules of paragraph 10.3, and, in any case, always make sure that the device is
switched off, without battery and with the pads disconnected.
The defibrillation pads are single-use, to be used on just one patient. Do not reuse the defibrillation pads; throw them
away after use and replace them with a new pair.
The defibrillation pads are not sterile or sterilisable.
The intense or prolonged administration of cardiopulmonary resuscitation with the defibrillation electrodes applied to
the patient can damage the electrodes. Replace them if they are damaged due to use or handling.
Inappropriate maintenance may damage SMARTY Saver Plus or cause it to malfunction. Comply with what is described
in this User Manual.
Use the non-rechargeable batteries SMT-C14031 of A.M.I. Italia S.r.l. by the indicated expiry date.
Remove the batteries from the device only if it has been off for at least 5 seconds. Otherwise, the device and the
batteries may be damaged.
SMARTY Saver Plus, its parts and accessories are manufactured non-sterile and non-sterilisable.
Do not expose SMARTY Saver Plus, its parts or accessories to direct light or high temperatures.
In case of failed connection or communication with the compatible Q-CPR module (SMT-C14034), SMARTY Saver Plus
will emit, in any case, the audio signal as metronome for the compressions, to help the operator in their operations.
All products, product data and specifications are subject to modification to improve their reliability, functionality,
design or other aspects.