SEMI-AUTOMATIC
26
7.6 Defibrillable rhythm
If a Ventricular Fibrillation or Tachycardia is detected, SMARTY Saver Plus will inform the operator with the following
commands:
Illuminated Icons/Buttons
Icon
“do not touch the patient”
Lit without blinking
Keep your distance, charging
Push the blinking red button
To deliver the shock, push the shock button within 15 seconds from the voice message, making sure that no one is
touching the patient.
If the shock button is not pushed within 15 seconds from the shock warning, SMARTY Saver Plus will disarm
automatically.
At this stage, the shock button will no longer blink and the device will inform the operator with the following voice
messages:
The shock button was not pushed
If, instead, the shock button is pushed within the 15 seconds, SMARTY Saver Plus will make the defibrillation shock.
At this stage, the shock button will no longer blink and the device will inform the operator with the following voice
messages:
SMARTY Saver Plus delivers the shock using the BTE waveform with automatic compensation of the patient’s thoracic
impedance. The value of the impedance detected must range between 20 and 200 Ohm; if the value detected is outside
of this range, the device will ask for the pads to be placed once more.
The shock protocol of SMARTY Saver Plus is incremental, i.e. the energy delivered to the patient varies incrementally
based on the number of shocks performed:
• first shock, energy 150J
• subsequent shocks at 200J.
This protocol is preset and cannot be modified by the user; it can be modified and customised solely and exclusively by
A.M.I. Italia S.r.l. at the express request of the client (endorsed by a competent body).