SEMI-AUTOMATIC
38
10.9 Patient analysis system
It determines patient impedance and assesses the ECG rhythm and the quality of the signal, to
determine whether administering the shock is appropriate or not.
≥4 seconds (with new, fully charged battery) in compliance with the standards IEC/EN 60601-2-4
97% in compliance with the standards IEC/EN 60601-2-4
99% in compliance with the standards IEC/EN 60601-2-4
If used on a patient with the characteristics listed in the use criteria, the SMARTY Saver Plus
defibrillator is designed to suggest a defibrillating shock when it detects the right impedance
and when the following situations occur:
Ventricular Fibrillation peak-to-peak amplitude at least 200µVolts
Ventricular tachycardia with heartbeat frequency min. 180 bpm and peak-to-peak amplitude at
least 200µVolts (including ventricular flutters and polymorphic Ventricular tachycardia)
Non-defibrillable rhythms
SMARTY Saver Plus is designed not to suggest shocks with all other rhythms, including: normal
sinus rhythm, moderate ventricular fibrillation (<200 μVolts), some slow ventricular tachycardias
and asystole.
10.10 ECG Analysis operation
Rhythm to be defibrillated
Ventricular Fibrillation (VF)
Rhythm to be defibrillated
Ventricular Tachycardia (VT, bpm >140)
Rhythm not to be defibrillated
Normal sinus rhythm
Rhythm not to be defibrillated
Asystole
Non-treatable rhythm
Generic AF, SVT, PVC
Positive predictive values
10.11 Defibrillator battery
Li-MnO₂ (Lithium - manganese dioxide) single-use, non-rechargeable
200 full rescue cycles (shocks + CPR) at 200J.
Duration in Stand-by mode
Estimated 3 (three) years supposing a battery activation test and daily self-tests,
without the AED being switched on
(at environmental conditions of temperature 20°C and humidity 45% without condensation)
*Perfo*Performance referred to a new and fully charged battery stored at a constant temperature of 20°C and relative humidity of
45 %4 45%without condensation
10.12 Internal back-up energy source
Electric Double Layer Capacitor (Super-Cap)
Preserving configuration data (date/time, etc.)
30 minutes (with AED battery inserted at least 15 minutes earlier)