SEMI-AUTOMATIC
39
10.13 Defibrillation pads
SMT-C2001 Pre-connected universal plates
SMT-C2002 Face-to-face pre-connected universal plates
Cable and connector external to the bag
Adult age >8 years or weight > 25Kg
Child age 1 - 8 years or weight < 25Kg
Number of shocks
tolerated
50 shocks at 360J (please refer to the specific user manual)
Medical FOAM, thickness 1 mm
Adhesive low-impedance conducting gel
Safety shock-proof connector
10.14 Timing of Shock cycles
Charge time performance in compliance with 60601-2-4 (201.101)
The maximum time between the start of the ECG Analysis and the completion of maximum
energy charge
The maximum time from when the AED is switched on until the completion of maximum
energy charge
2400.00 (MHz); 2440.00(MHz); 2485.00(MHz)
• Communication with the external CPR Quality SMT-C14034 sensor
• Internal 57600 baud Serial Port