SEMI-AUTOMATIC
21
6 Self-test
SMARTY Saver Plus was designed to be a totally safe device, always ready to use and able to automatically and
constantly check the proper operation of its parts, reducing maintenance operations by the user to a minimum.
In fact, SMARTY Saver Plus performs three types of self-test:
Activation: On insertion of the battery
Automatic: In stand-by mode, daily/monthly/bi-annually
On: On the device being switched on
The outcome of the control test can be viewed through the control LEDs (green and/or red) that make it possible to see,
at any time, when the device is switched off (stand-by mode), the operating status of the device and its main accessories.
SMARTY Saver Plus performs the operation tests only when the battery is installed; we, therefore, recommend not to
remove the battery from the device, except only for the time needed to replace it.
6.1 ACTIVATION test
On each insertion of the battery, the device will perform the ACTIVATION diagnostic test; this self-test results in a small
consumption of energy, as it involves all components of the device and also requires a manual intervention by the
operator, who must:
Insert the battery in the device
If the battery has been inserted properly, SMARTY Saver Plus will automatically switch on emitting an acoustic signal,
and the on button will turn green, while the control LED will switch off.
The device will emit voice messages suggesting the operation to be followed.
“Device test”
“Push the blinking red button”
The shock button will start blinking.
The operator will have a maximum of 60 seconds* to push the shock button.
Push the shock button to launch the ACTIVATION test*
When the shock button is pushed, it will stop blinking and the
device will launch the activation test.
If the test ends with an error, the device switches off automatically and the red control LED will blink approximately
every 6 seconds. If, instead, the device activation test concludes without errors, the device will emit the voice message
"place the Pads"; the two red LEDs of the "place the PADs" indicator on the keypad will blink.
Switch off the device
If immediate use is not required, switch off SMARTY Saver Plus and leave the battery inserted to guarantee the
performance of periodic diagnostic self-tests (see Paragraph 6.2)
*If the shock button is not pushed within 60 seconds, SMARTY Saver Plus detects an error and switches off; disconnect and reconnect
the battery to repeat the activation test and push the shock button within the time limit, to complete the test.
If, instead, the shock button was pushed but continues to blink, the shock button is not working properly.
In such case, contact the authorised technical support centre.