EasyManua.ls Logo

Smarty Saver Plus - CPR Quality Module

Default Icon
47 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SEMI-AUTOMATIC
20
5.3 CPR Quality module
SMARTY Saver Plus assists the operator in properly performing cardiac massage, during Cardio-Pulmonary
Resuscitation, thanks to the external CPR Quality sensor (SMT-C14034).
This external device (class I) is, in fact, able to measure the depth and the frequency of the compressions performed
and to send this feedback to the SMARTY Saver Plus device via Bluetooth according to the international AHA / ERC
guidelines, for both adult and paediatric patients, as indicated in the table below:
COMPRESSIONS
ADULT/ADOLESCENT
CHILD
FREQUENCY
100-120/min
DEPTH
At least 2 inches (5 cm) but no more
than 2.4 inches (6 cm)
At least one third of the chest
Approximately 2 inches (5 cm)
Thanks to the CPR Quality module, the operator can check:
the correctness of the depth of the compressions they are performing, through the LED bar on the
defibrillator’s keyboard (see figure CPR Feedback)
the correct frequency/rhythm of compressions through audio signals emitted by the AED.
Compatible external Q-CPR module (SMT-C14034)
switch the module on by pushing the side on key
place it on the patient’s chest at the time of CPR
Perform the compressions, controlling their correctness
through the LED bar on the device’s keyboard and with the
use of the voice instructions emitted by the AED
Progressive LED scale:
red LED on: insufficient or excessive pressure
red and orange LED on: not enough pressure
red, orange and yellow LED on: almost sufficient pressure
red, orange, yellow and green LED on: correct pressure
Note: For more information on the external Q-CPR sensor (SMT-C14034), please refer to the specific user manual.
CPR Feedback
Q-CPR module

Table of Contents

Related product manuals