9.3 Cleaning
The structure of the SMARTY Saver Plus defibrillator, including the port for the connection of the defibrillation
electrodes, can be sanitised with the help of a soft cloth dampened with one of the detergent solutions listed below:
a) Isopropyl alcohol (70% solution)
b) Soapy water
c) Bleach (30 ml per litre of water)
d) Detergents containing ammonia
e) Detergents containing glutaraldehyde
f) Hydrogen peroxide
Do not immerse SMARTY Saver Plus in any liquids.
Do not use abrasive materials or detergents, strong solvents, such as acetone or acetone-based detergents,
and enzymatic detergents.
Do not sterilise SMARTY Saver Plus or its accessories.
9.4 Storage
SMARTY Saver Plus must be placed in a place that complies with the environmental and safety conditions of the table
below and at the temperature and humidity indicated in paragraph 10.2
The device must be stored with the battery always inserted, to allow periodic self-testing.
So that the device is easy to find in case it is needed, place it where it will be easily accessible and turned around so that
the control LEDs are visible.
Do not use, install or store
SMARTY Saver Plus in
temperature or humidity
conditions outside of the
ranges mentioned in this
user manual.
Do not install or store
SMARTY Saver Plus in
areas directly exposed to
sunlight
Do not install or store
SMARTY Saver Plus in
areas subject to significant
fluctuations of
temperature or humidity
Do not install or store
SMARTY Saver Plus near
sources of heat
Do not use, install or store
SMARTY Saver Plus in
places subject to strong
vibrations
Do not use, install or store
SMARTY Saver Plus in
spaces with a high
concentration of
anaesthetic or inflammable
gases
Do not install or store
SMARTY Saver Plus in
areas with high
concentrations of dust
Tampering with SMARTY
Saver Plus can only and
exclusively be done by
A.M.I. Italia or personnel
authorised thereby.