SEMI-AUTOMATIC
35
10.4 Table of Alarms
Priority
Cause
Visual signal
HIGH
Device ready to administer shock
Blinking shock button
HIGH
Battery empty (capacity < 1%)
Blinking control LED
10.5 Controls and indicators
Category
Rated specifications
Buttons
ON/OFF: switches the device on and off
Shock: delivery of the defibrillation shock
“Adult” selection
“Paediatric” selection
Visual Indicators
Device status control LED (2 LEDs: red and green)
Defibrillation pad placement LED (2 red LEDs)
Do not touch the patient LED (2 red LEDs)
You can touch the patient LED (1 green LED)
Adult patient LED (1 green LED)
Paediatric patient LED (1 green LED)
ON/OFF button LED (2 green LEDs)
Shock button LED (8 red LEDs)
Bar of 2 red + 2 orange + 2 yellow + 2 green LEDs (for Q-CPR)
Bluetooth connection with Q-CPR LED (1 green LED)
Audio Indicators
Audio messages for instructions during use
Warning and hazard acoustic signals
Speaker
Preset volume (Emissions compliant with IEC/EN 60601-2-4 point 6.1)
Variation min. 20% max 100% (60 dBA to 80dBA ±3 dBA)
Microphone
Automatically activated recording on device switching on
10.6 Data memory
Category
Rated specifications
External memory (optional)
µSD/SDHC-type Memory Card up to 32 GB (max)
Stored data
AED1LOG.txt
Daily self-tests, Errors detected, Device use data,
Device information
AEDFILE.aed
Rescue data, Environmental voices and sounds, Rescue ECG trace,
Analysed and detected patient vital parameters
Data display
Through PC Saver View Express Software (Microsoft Windows compatible)

Table of Contents