MPI-540 ● MPI-540-PV – BRIEF USER MANUAL
8.3 Other technical data
a) type of insulation acc. to EN 61010-1 and IEC 61557 ........................................................... double
b) measurement category acc.to EN 61010-2-030 ........... IV 300 V, III 500 V, II 1000 V DC
c) ingress protection acc. to EN 60529 .......................... IP51 (terminals sealed with silicone stopper)
d) power supply of the meter ................................................................. Li-Ion 11,1 V 3,4 Ah 37,7 Wh
e) parameters of AC adapter for the battery charge .................................................... 12 V DC / 2,5 A
...................................................................................... 100 V…240 V, 50 Hz…60 Hz (power grid)
f) dimensions ..........................................................................................288 mm x 223 mm x 75 mm
g) weight of the meter with batteries ............................................................................. approx. 2,5 kg
h) storage temperature .................................................................................................. -20°C...+60C
i) working temperature ..................................................................................................... 0°C...+45C
j) temperature range suitable for initiating battery charging.......................................... +10C...+40C
k) temperatures at which loading is interrupted.................................................... <+5 C and ≥ +50C
l) humidity ......................................................................................................................... 20%...90%
m) reference temperature ................................................................................................. +23°C ± 2C
n) reference humidity ......................................................................................................... 40%...60%
o) altitude (above sea level) ................................................................................................... <2000 m
p) time settings for automatic shut-down when idle (Auto-OFF) ..................... 2 min, 5 min or disabled
q) number of measurements Z or RCD (for charged batteries) .. >3000 (6 measurements per minute)
r) number of measurements RISO or R (for charged batteries) .................................................. >1000
s) time of recording (for charged batteries) .................................................................................. 16 h
t) display ............................................................................................. color LCD TFT with polytouch,
............................................................................................................................. 800 x 480 pixels,
...................................................................................................................................... diagonal 7”
u) memory of measurement results ...................................................................................... unlimited
v) recorder memory .............................................................................................................. unlimited
w) data transmission .................................................................................................................... USB
x) quality standard ..............................................................................................................................
.............. design, construction and manufacturing are ISO 9001, ISO 14001, ISO 45001 compliant
y) the device meets the requirements of IEC 61557 standard
z) the product meets EMC requirements (immunity for industrial environment) according to the fol-
lowing standards............................................................................. EN 61326-1 and EN 61326-2-2
EN 55022 Compliance statement
MPI-540 / MPI-540-PV is a class A product. In a domestic environment this product
may cause radio interference in which case the user may be required to take ade-
quate measures (e.g. increasing the distance between affected products).