Page #64 background image
4 Customizing Your System
56 Basic User Manual
Exam Preset
Comment
Bodymark
Measure
General
DICOM
Help
Load Default
About
Apply
Exit
SystemSettings
Report
Peripheral
Service List
DICOM Service Local Setting
Add Edit Delete Ping Verify
Service Type
StoreSCP DicomStore AE 192.168.254.104 104
Service Name AE Title IP Address Port
Figure 4-18 DICOM Service Settings Screen
■  Storage Server Settings
Add Dicom Service
Service Type Storage
Service Name
AE Title
IP Address
Port Number
Storage Commitment Server
Ping
Verify
ResetCancelSave
Item Description
Service Name Set the server name of the DICOM storage server.
AE Title Set the Application Entity (AE) title of the DICOM storage
server.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sonoscape E1 and is the answer not in the manual?

Sonoscape E1 Specifications

General IconGeneral
TypePortable Ultrasound System
Display15-inch LCD
Transducer Ports2
Cine LoopYes
Imaging ModesColor Doppler, Power Doppler
ApplicationsObstetrics, Gynecology, Vascular, Musculoskeletal
Probe Connectors2
PortsVGA, USB, Ethernet
Supported ProbesConvex, Linear, Phased Array