Chapter 6: Safety 91
Safety
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 9: Transducer Model: C60x Operating Mode: 2D
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1A
aprt
>1
Global Maximum Index Value
1.0 (a) — — — (b)
Associated Acoustic
Parameter
p
r.3
(MPa)
1.69
W
0
(mW)
#— —#
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
z
bp
(cm)
—
z
sp
(cm)
4.7 —
d
eq
(z
sp
)(cm)
—
f
c
(MHz)
2.84 # — — — #
Dim of A
aprt
X(cm)
#— — — #
Y(cm)
#— — — #
Other Information
PD (μsec)
0.579
PRF (Hz)
5440
p
r
@PII
max
(MPa)
2.679
d
eq
@Pll
max
(cm)
—
Focal Length FL
x
(cm)
#— — #
FL
y
(cm)
#— — #
I
PA.3
@MI
max
(W/cm
2
)
197.7
Operating
Control
Conditions
Control 1: Exam Type
Abd/OB
Control 2: Optimization
Any
Control 3: Depth
11/
13 cm
Control 4: THI
On
Control 5: MB (Multi Beam)
On