90
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 8: Transducer Model: C11x Operating Mode: PW Doppler
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1A
aprt
>1
Global Maximum Index Value
(a) — 1.0 — 1.7 1.8
Associated Acoustic
Parameter
p
r.3
(MPa)
#
W
0
(mW)
— 46.0 24.9 25.4
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
z
bp
(cm)
—
z
sp
(cm)
# 1.06
d
eq
(z
sp
)(cm)
0.24
f
c
(MHz)
#—4.36 — 4.37 4.36
Dim of A
aprt
X(cm)
— 1.76 — 0.28 0.20
Y(cm)
— 0.50 — 0.50 0.50
Other Information
PD (μsec)
#
PRF (Hz)
#
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
0.23
Focal Length FL
x
(cm)
—6.37 — 0.77
FL
y
(cm)
—4.40 — 4.40
I
PA .3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Exam Type
Any Any Any
Control 2: Sample Volume
2 mm 1 mm 1 mm
Control 3: PRF
3906 10417 20833
Control 4: Sample Volume Position
Zone 7 Zone 1 Zone 0