Chapter 6: Safety 89
Safety
(a) This index is not required for this operating mode; value is <1.
(b) This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
— Data are not applicable for this transducer/mode.
Table 7: Transducer Model: C11x Operating Mode: CPD/Color
Index Label MI
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
≤1A
aprt
>1
Global Maximum Index Value
(a) (a) — — — 1.0
Associated Acoustic
Parameter
p
r.3
(MPa)
#
W
0
(mW)
#— —38.8
min of [W
.3
(z
1
),I
TA.3
(z
1
)] (mW)
—
z
1
(cm)
—
z
bp
(cm)
—
z
sp
(cm)
# —
d
eq
(z
sp
)(cm)
—
f
c
(MHz)
## — — — 4.37
Dim of A
aprt
X(cm)
# — — — 1.12
Y(cm)
# — — — 0.50
Other Information
PD (μsec)
#
PRF (Hz)
#
p
r
@PII
max
(MPa)
#
d
eq
@Pll
max
(cm)
—
Focal Length FL
x
(cm)
#— — 4.29
FL
y
(cm)
#— — 4.40
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Mode
Any
Control 2: Exam Type
Abd
Control 3: PRF
3676
Control 4: Optimization/Depth
Low/5.1
Control 5: Color Box
Position/ Size
Top/
Short &
Narrow