9 Technical specifications
208396 rev 2 SOREDEX 49
9.2 System requirements and connections
Minimum requirements for the PC/laptop, network adapter and network switch
PC / laptop network switch Class I or Class II according IEC 60950
Network connection settings
10/100Mbs LAN
UDP/IP protocol traffic allowed
Traffic to UDP port 10000 allowed (unit UDP port)
UDP broadcast traffic allowed
CAT6 Ethernet cable
DHCP server is recommended but not necessary
Use
Use antivirus software.
Use firewall.
When LAN configuration is changed or devices are
added/removed, it may affect existing devices in
the LAN. Therefore keep in mind that correct oper-
ation of the imaging system needs to be checked
after changes are made.
When adding new devices to LAN, make sure they
all have unique IP address, otherwise they may
cause communication problems with existing LAN
devices. Place unit and PC with imaging applica-
tion software to same subnet in LAN.
NOTE! Image is not transferred from unit to PC imaging application software in case of
connection lost during image processing. Image is stored in unit memory until it has been
transferred to PC. Unit cannot be turned off in that case. When network is operational again,
image is automatically transferred to imaging application software. Do not disconnect unit PSU
adapter before network is operational and image has been transferred to imaging application
software.
For more details of the hardware requirements running the imaging application software please
refer to the user manual of it.