9 Technical specifications
52 SOREDEX 208396 rev 2
9.5 Electromagnetic Compatibility (EMC)
tables
Note! Medical electrical equipment needs special precau-
tions regarding EMC and needs to be installed according to
EMC information.
Note! Mobile RF communications equipment can effect
medical electrical equipment
Guidance and manufacturer’s declaration - electromagnetic emissions
The DXR-60 is intended for use in the electromagnetic environment specified below. The
customer or the user of the DXR-60 should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic environment - guidance
RF emissions
CISPR 11E
Group 1 The DXR-60 uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B The DXR-60 is suitable for use in all establishments,
including domestic establishments and those direct-
ly connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies