10 Symbols and labeling
208396 rev 2 SOREDEX 59
10.3 Warnings and precautions
THE UNIT IS A CLASS 1 LASER PRODUCT
Note! When covers are removed the unit is a class 3B laser
product – avoid exposure to the laser beam.
Use of controls or adjustments or performance of
procedures other than those specified herein may result in
hazardous laser radiation exposure or other non-
compliance.
When handling imaging plates, protective covers
and hygiene bags always take the appropriate hy-
giene measures and precautions to prevent cross
contamination. New protective cover must be used
on every use.
The imaging plates are harmful if swallowed.
Do not move or knock the unit when it is processing
an imaging plate.
This unit must only be used to process image
plates supplied by the manufacturer and must not
be used for any other purpose.
NEVER use imaging plates, protective covers or
hygiene bags from other manufacturers.
This unit, or its accessories, must not be modified,
altered or remanufactured in any way.
Only the manufacturer’s authorized service per-
sonnel are authorized to carry out maintenance
and repair of the unit. There are no user servicea-
ble parts inside the unit.
Unit is not suitable for use in the presence of flam-
mable anaesthetic mixture with air or with oxygen
or nitrous oxide.
In order to maintain safe and correct functioning of
the unit, only the power supply unit (PSU) delivered
with the unit or distributed by authorized dealers.
Please refer to the unit technical specifications for
a list of the PSUs.