10 Symbols and labeling
60 SOREDEX 208396 rev 2
■ For ethernet connections, use an unshielded CAT6
RJ-45 LAN cable, so that multiple chassis must not
be connected! The PC / Ethernet switch to which
unit is connected to must be class I or class II ac-
cording IEC 60950. After installation check that the
IEC 60601-1 leakage current levels are not ex-
ceeded.
■ If the PC/ Ethernet switch to which the device is
connected to is used in the patient environment it
should be approved appropriately and meet the re-
quirements of 60601-1 standard.
■ The PC and any other external device(s) connect-
ed to the system outside the patient area must
meet the IEC 60950 standard (minimum require-
ment). Devices that do not meet the IEC 60950
standard must not be connected to the system as
they may pose a threat to operational safety.
■ The PC and any other external devices shall not be
connected to an extension cable.
■ Multiple extension cables shall not be used.
■ If this device will be used with 3rd party imaging ap-
plication software not supplied by the manufactur-
er, the 3rd party imaging application software must
comply with all local laws on patient information
software. This includes, for example, the Medical
Device Directive 93/42/EEC and/or FDA if applica-
ble.
■ Do not position the PC where it could be splashed
with liquids.
■ Clean the PC in accordance with the manufactur-
er’s instructions.
■ Image is not transferred from unit to PC imaging
application software in case of connection lost dur-
ing image processing. Image is stored in unit mem-
ory until it has been transferred to PC. Unit cannot
be turned off in that case. When network is opera-
tional again, image is automatically transferred to
imaging application software. Do not disconnect
unit PSU adapter before network is operational and
image has been transferred to imaging application
software.