9 Technical specifications
208396 rev 2 SOREDEX 53
Guidance and manufacturer’s declaration – electromagnetic immunity
The DXR-60 is intended for use in the electromagnetic environment specified below. The
customer or the user of the DXR-60 should assure that it is used in such an environment.
Immunity Test IEC 60601
Test Level
Compliance
Level
Electromagnetic Environment
- guidance
Electrostatic
discharge
(ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete
or ceramic tile. If floors are cov-
ered with synthetic material, the
relative humidity should be at
least 30%.
Electrical fast
transients/
bursts
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential
mode
line to line
± 1 kV differen-
tial mode
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short interrup-
tions and volt-
age variations
on power sup-
ply lines
IEC 61000-4-
11
< 5% U
T
(> 95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in U
T
)
for 5 sec
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(> 95% dip in
U
T
)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If user of
the DXR-60 requires continued
operation during power mains in-
terruptions, it is recommended
that the DXR-60 be powered
from an uninterruptible power
supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic field
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment.
NOTE: U
T
is the AC mains voltage prior to application of the test level.