40 SPECTROTEST —
Fault Possible cause Remedy
The unit does not
generate sparks in the
arc or spark mode.
The measuring window of
the software has a "not
active" status in Windows.
Press "Esc" or click the
measuring window with the left
"mouse button" of the trackball.
The electrode gap is too
small.
Clean the electrode and re-
the gap using the spacer.
The sample probe start
button is defective.
Please contact your service
center.
The plasma generator
parameters of the program
that is used were modified
or deleted.
Have the former settings
restored by qualified
personnel.
Please contact your service
center.
The plasma generator is
defective.
Please contact your service
center.
The actual values are
below the stored limit
values for the reference
elements (error
message).
The adapter is reversed in
the arc mode.
Turn the adapter so that one
opening faces the fiber-optics
window.
The spark (arc) adapter is
not locked in properly up
to the stop.
Check whether the adapter is
properly seated and correct it if
necessary.
The electrode gap is not
correct.
Adjust the electrode gap using
the spacer.
The samples have not
been properly prepared.
See the information given under
“No correct focal spot in the spark
mode”.
The material was not
measured with a program
associated with the base
(e.g. Co sample with a Fe
base program).
Go to the suitable global program
if available (e.g. to Co-01)