Safety Information and Warnings
RoboSep™-S (Catalog #21000) must be operated by trained professionals and is intended for research use
only. Not intended for human or animal diagnostic or therapeutic uses.
RoboSep™-S should be placed on a clean, flat surface free of any foreign objects or debris and insulated
from excessive vibration. There should be at least 13 mm of clearance on each side of the instrument to
ensure adequate cooling.
Always connect the power supply to a 3-prong, grounded AC outlet rated 2A 100 - 240 V using the AC power
cord provided with RoboSep™-S. Do not use an adapter to a two-terminal outlet. Before plugging the
instrument in, be sure that two FB 4 A 250 V~ fuses are installed.
All electrical devices pose a risk of electric shock. To reduce this risk, do not open any covers that are
fastened with screws. While RoboSep™-S is designed to withstand spills, do not allow fluids to enter the
interior of the instrument. In the event of such a spill, switch off RoboSep™-S and disconnect the power cable
before cleaning (section 6.2).
Operate RoboSep™-S at 15 - 30°C, in an environment with a relative humidity of 20 - 85%. RoboSep™-S
should be used indoors only. RoboSep™-S does not require placement in a biohazard safety cabinet and is
not specified for use inside an incubator or cold room. Protect the instrument against dust and moisture. Do
not exert excess pressure or use sharp objects on the touchscreen. It is not recommended to use UV light
lamps for prolonged periods of time, as exposure to UV light will result in discoloring of external RoboSep™-S
components over time.
EasySep™ magnets on the RoboSep™-S carousel generate strong magnetic fields. Keep away from
pacemakers, watches, or other objects that respond to magnetic fields.
RoboSep™-S weighs 22 kg (48.5 lb). Exercise caution when moving RoboSep™-S. It is recommended that
two people are present to lift or move the instrument safely.
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and (2) this device must accept any interference received,
including interference that may cause undesired operation.
In case of malfunction, call STEMCELL Technologies Inc. for service. Servicing performed by anyone other
than an authorized STEMCELL Technologies representative will void the warranty associated with
RoboSep™-S. There are no user-serviceable parts inside the instrument and the outer casing should never
be removed.
RoboSep™-S is manufactured and distributed by STEMCELL Technologies Inc., 1618 Station Street,
Vancouver BC, Canada V6A 1B6. Call 1-604-877-0713 or toll-free 1-800-667-0322 within North America.
STEMCELL Technologies is not responsible for any injury or damage caused by the use of this instrument for
purposes other than those for which it is intended, or by modifications of the instrument or software not
performed by an authorized STEMCELL Technologies representative.
RoboSep™-S Software
This manual is applicable to RoboSep™-S software version 1.3.0.0. For older software versions, contact us at
techsupport@stemcell.com for the applicable manual.