AbraPol-10
Instruction Manual
25
Different preparation steps can be programmed individually in
regard of time, speed of disc, water, type and dosing of lubricant
and suspension. These steps can be combined in a number of
different ways so that different materials can be prepared without
having to repeat the programming.
Move the nozzle block to the side.
Place the disc on the turntable and move it until the three
pins engage with the holes in the turntable.
Place the specimen holder under the quick coupling.
Press and hold the black flange of the column down with your
thumbs while guiding the pressure tap of the specimen holder
into the coupling.
Release the black flange.
Turn the specimen holder until the three pins engage with the
corresponding holes.
The position of the specimen holder has to be adjusted correctly in
relation to the preparation disc to get the best possible
preparation results.
Loosen the two handles placed on each side of the specimen
holder motor.
The specimen holder motor can now be moved manually to the
sides. The eccentricity can be read on the scale placed on the
specimen holder motor.
When the correct position is found fasten the handles again.
Check that the splash ring is in place.
Press ESC to get to Main Menu.
Select Struers Methods in Main Menu by turning the knob.
Push the knob for entering Struers Methods.
Turn the knob to select a Struers Method.
View Method by pushing the knob.
Start the preparation by pressing the two double start buttons
simultaneously.
Keep the buttons pressed until the sample motor starts.
Programming a Preparation
Step
Mounting a Preparation Disc
(300 or 350 mm)
Inserting the Specimen Holder
Adjusting the Specimen Holder
Position
Starting the Preparation
Process (Struers Method)